Figure 16-2 SideStream Module
1.
The Sidestream module can be used for intubated and non-intubated patient.
2.
Insert LEMO interface to socket( CO
2
) of the monitor
3.
Connect the sample cell to the module,the sampling pump automatically turns on.Perform sample cell zero.
4.
Place the sampling line on the patient. Removal the sample cell automatically turns sampling pump off.
5.
Capnogram displayed in less than 20 seconds, at an ambient temperature of 25°C, full specifications within 2
minutes.
Mainstream
Figure 16-3 MainStream Module
1.
The mainstream module can only be used for intubated patient.
2.
Set the "WORK MODE" to "MEASUREMENT"in the submenu "OTHER SET" of "CO
2
MENU".
3.
Connect the module to the monitor.
4.
Place the module above on the adapter.
5.
Perform airway adapter zero.
6.
Connect the airway adapter into the patient's airway.
7.
Capnogram, displayed in less than 15 seconds at an ambient temperature of 25ºC, full specifications within 2
minutes.
Warning
Do not use the accessory if the packaging or the internal accessory is damaged. Return it to the manufacturer.
The sidestream sampling line is disposable. It should not be disinfected for reuse or cross-used by different
patients.
In the long-term use, dust or other substances may lower the air permeability of the filter material in the sample
cell and may block the airway. In this situation, the sampling line must be replaced.
No routine user calibration required. For mainstream,an airway adapter zero is required when changing to a
different style of airway adapter.