8.7.7 Illumination is switched off at setting a higher current
Description:
After setting a higher current, e.g. from 400 mA to 1500 mA, the controller and the
module are switched off.
Possible reason:
The power supply is too weak for the programmed current.
Solutions:
Check the power supply for correct output range. The XLC4 needs approximately 3.5
amps per channel at full operation. For more information, see section 6.4.2.
8.7.8 Illumination is switched off unexpectedly at regular operation
Description:
The module is switched off unexpectedly at regular operation. The module is quite hot.
Possible reason:
The module gets to hot and has been switched off by the controller.
Solutions:
Check for error messages
with the “ER” command. If it is the temperature error (205), let
the module cool down and start it again. Check for correct heat sink selection regarding
to the current setting. Check for correct operation of the fans.