8.6 Firmware update XLC4
Wrong operations at the update process might cause a loss of data at the update
process. In this case the update process might be interrupted and could not be
finished. The XLC4 controller must be returned to the manufacturer in this case.
It is strongly recommended to contact Chromasens support before starting a
firmware update.
The firmware of the XLC4 LED controller is able to be updated in the field by the user.
Notes
:
Saved settings in the XLC4 controller get lost at the update. It is recommended to
check the setting and to take notes of them.
Requirements
:
1. Connection to the XLC4 controller with RS232 port. Ethernet or USB is not applicable.
It is required to use COM port number 1 to 9
because
the tool “PIC32UBL.exe” only
supports this ports.
2. Software-
Tool „PIC32UBL.exe“ – Please ask Chromasens support
(email:
3. XLC4 Commander
Steps
:
1. Connect the XLC4 LED controller with RS232 to the PC.
2. Start the XLC4 commander and connect it with the XLC4 LED controller.
3. Type the command
UP
into the command line and click
Send
. This command erases
the old firmware: