(2) Since the window indicating the voltage value for inputting is displayed, input its value to this
recorder.
(3) Adjust the zero point.
(Example) For the adjustment of the ±2V range
Input the voltage of 0V with the DC standard voltage generator.
(4) After inputting the zero point for about 5 seconds, click the “Go” button.
(5) Adjust the span point.
(Example) For the adjustment of the ±2V range
Input the voltage of +2V with the DC standard voltage generator.
(6) After inputting the span point for about 5 seconds, click the “Go” button.
(7) After the adjustment of the span point, the screen is returned to the calibration screen for all ranges.
(8) Repeat from (1) to (6) for the adjustment of other ranges.