INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued)
Secure Therapy System to Module
Carefully place the Therapy System and
Module on one side. With a #1 Phillips
Screwdriver, install the four 4 mm x 20 mm
screws.
Tighten screws until the Module does not
move on the Therapy System.
Front Access Panel
With a #1 Phillips Screwdriver, remove the
screw retaining the existing Front Access
Panel.
Install Lanyard to the new Extended Front
Access Panel using the same screw.
Do not twist Ribbon Cable. If power is
applied to the system with misalignment of
pins or a twisted ribbon cable, the controlling
electronics in the Module will be destroyed
and possible damage to the System's internal
components could occur.
Set Therapy System onto Module
Set the Therapy System on the Module.
Make certain the Feet of the Therapy
System are resting in the Module Indents.