Intelect® Advanced Therapy System
SAFETY PRECAUTIONS
5
Stimulation should not be applied over the anterior neck or mouth.
Severe spasm of the laryngeal and pharyngeal muscles may occur and
the contractions may be strong enough to close the airway or cause
difficulty in breathing.
Stimulation should not be applied transthoracically in that the
introduction of electrical current into the heart may cause cardiac
arrhythmia.
Stimulation should not be applied over swollen, infected, and
inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis,
varicose veins, etc.
Stimulation should not be applied over, or in proximity to, cancerous
lesions.
Output current density is related to electrode size. Improper application
may result in patient injury. If any question arises as to
the proper electrode size, consult a licensed practitioner prior to therapy
session.
•
•
•
WARNINGS (continued)
Supplied by EME Services Ltd