Intelect® Advanced Therapy System
71
OPERATION
Refer to pages 35-38 for Electrotherapy
patient preparation, select electrodes, and
secure electrodes. Refer to page 39 for
Ultrasound patient preparation.
Press the Up and Down Arrow buttons to
highlight the desired treatment date and
time.
Press the View Treat. button.
Press Start New Treatment button.
EXISTING PATIENT DATA CARD USE (continued)
Starting a New Treatment from Patient
Data Card
VIEW TREAT.
BUTTON
START NEW
TREATMENT
BUTTON
Set intensity by rotating the Intensity Control
Knob to the prescribed level.
Set Intensity
INTENSITY
DISPLAYED
Clockwise-
Increases Intensity
Counterclockwise-
Decreases Intensity
Intensity Knob Rotation
ROTATE
INTENSITY
KNOB
Connect Patient Interrupt Switch to the
Therapy System. Give Patient Interrupt
Switch to patient and explain that pressing
the Red button once pauses the therapy
session.
Patient Interrupt Switch
CONNECT PATIENT
INTERRUPT SWITCH
PATIENT
INTERRUPT
SWITCH
If Patient Interrupt Switch is depressed, the
treatment will be paused and a message
will appear on the System screen.
Press any button to clear the message.
NOTE:
If the Patient Interrupt Switch is depressed
a second time, the message will clear from
the screen and the treatment will remain
paused.