Intelect® Advanced Therapy System
SPECIFICATIONS
24
It is a monophasic waveform with a phase duration of 2 ms and a
pause of 5 ms resulting in a frequency of approximately 143 Hz.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA
Polarity Reversal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . On or Off
With Polarity Reversal On, Polarity will change every 7.5 minutes.
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
A series of rectangular, monophasic pulses. The pulses surge to
maximum power, hold and then decrease before the pause. This
waveform is well suited for muscle strengthening.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA
Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.2-5.0 ms
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-60 Hz
Surges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1/min - 20/min
Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-57 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
Stimulus delivered by the TENS waveforms of this device, in
certain configurations, will deliver a charge of 25 microcoulombs
(µC) or greater per pulse and may be sufficient to cause
electrocution. Electrical current of this magnitude must not flow
through the thorax because it may cause a cardiac arrhythmia.
Träbert (Ultrareiz)
SURGED: Monophasic Rectangular
WAVEFORM SPECIFICATIONS (continued)