Intelect® Advanced Therapy System
57
OPERATION
COMBINATION
The Intelect Advanced Therapy System Combination modality allows the user to select and use ultrasound therapy in combination with
electrical muscle stimulation.
Combination therapy utilizes the Ultrasound modality in conjuction with High Voltage Pulsed Current (HVPC), IFC (4p), IFC (2p), Asymmetrical
Biphasic, Symmetrical Biphasic, or VMS
TM
to generate a therapeutic effect. In this mode of therapy, the Sound Head of the Ultrasound
Applicator becomes one half of the electrical circuit. An electrode attached to the Red (+) Lead Wire completes the circuit.
Press the Combination button on the Home
screen.
Refer to pages 35 through 38 to prepare
patient, select electrode, and securing
electrodes. Refer to page 39 for Ultrasound
patient preparation.
Connect the Black (-) Lead Wire from
Channel 2 to the electrode. Make certain
the Lead Wire is completely seated in the
electrode.
The Red (+) Lead Wire is not used. The
Ultrasound Applicator completes the
circuit for Combination Therapy.
COMBINATION
BUTTON
Select Modality
Prepare Patient
Press the Waveform Description button
to view text explaining the waveform
rationale.
Press the Back button to return to the
Treatment Review screen.
View Application Description
APPLICATION
DESCRIPTION
BUTTON
BACK
BUTTON
Supplied by EME Services Ltd