Intelect® Advanced Therapy System
91
OPERATION
Give Patient Interrupt Switch to patient
and explain that pressing the Red button
once pauses the therapy session.
The first waveform/current in the Sequence
should be highlighted. Set intensity by
rotating the Intensity Control Knob to the
prescribed level.
Set Sequence Intensity
PRESSING BUTTON ONCE PAUSES SESSION
INTENSITY DISPLAYED
Clockwise-
Increases Intensity
Counterclockwise-
Decreases Intensity
Intensity Knob Rotation
If Patient Interrupt Switch is depressed, the
treatment will be paused and a message
will appear on the System screen.
Press any button to clear the message.
NOTE:
If the Patient Interrupt Switch is depressed
a second time, the message will clear from
the screen and the treatment will remain
paused.
CLINICAL RESOURCES LIBRARY USING SEQUENCES (continued)
Press the Down Arrow button until the
second waveform/current in the sequence is
Highlighted.
Set intensity by rotating the Intensity Control
Knob to the prescribed level.
DOWN
ARROW
BUTTON
Repeat for the third waveform/current in the
Sequence.
Supplied by EME Services Ltd