Intelect® Advanced Therapy System
109
TROUBLESHOOTING
ERROR CODES (continued)
Code
Number
Type
Message
Probable Cause
Possible Remedies
118
Warning
Attempting to save additional User Protocols or Sequences after
system memory has reached the maximum allowed (200).
Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy System User
Manual for instructions.
119
120
121
122
Warning
Warning
Warning
Warning
1.
Attempted to read a treatment from Patient Data Card that is
not a valid treatment for the system
2.
Attempted to use a Non-Patient Data Card.
3.
No Patient Data Card inserted into system port.
4.
Unknown type of smart card inserted into system.
A.
Use a Patient Data Card with proper treatment data for the system.
B.
Properly insert a Patient Data Card.
C.
Insert a known good Patient Data Card.
D.
If problem persists, insert a known good Patient Data Card. If problem continues, contact
dealer or Chattanooga Group for service.
123
Warning
Patient Data Card is full.
Erase Patient Data Card. Refer to Therapy System User Manual for instructions.
124
Warning
Patient Treatment Data already saved.
A.
Cannot save same data again on Patient Data Card.
B.
Use a new Patient Data Card to resave data.
C.
Erase Patient Data Card and resave treatment data.
125
Warning
Multimedia Card (MMC) not in system port.
A.
Properly insert the MMC card into the system port.
B.
Insert a known good MMC Card. If problem continues, contact dealer or Chattanooga Group
for service.
126
Warning
No valid channels are available for attempted treatment.
A.
Complete existing treatment before attempting to start another.
B.
Reset Therapy System by turning main power switch Off and On.
127
128
Warning
Warning
1.
No sEMG Channels are available for treatment.
2.
No sEMG Module installed or detected by system.
A.
Wait until current treatment is complete.
B.
Reset Therapy System by turning main power switch Off and On.
C.
Make certain sEMG Module is properly installed. Refer to sEMG Module User Manual for
installation instructions.
D.
Replace sEMG Module with known good sEMG Module.
E.
If problem continues, contact dealer or Chattanooga Group for service.