Typical Applications
25
→
【
G/n
】
in input box of Marker 2. Data of these 2 frequency points will be displayed on the screen.
Figure 3-23 Amplifier Test Displayed in Four Windows
Measurement of 1dB compression point of amplifier:
you may connect 20dB attenuator first at
output end of amplifier and reset analyzer. Press
【
Sweep
】
key on front panel to call up
Sweep
soft key
menu displayed as shown in the left side of the following picture and then press
[Sweep Type]
. Soft key
menu will be as the middle menu shown in the picture below. Select
[Power Sweep]
.
Figure 3-24 Set up Power Sweep
Press
[Power]
in
Sweep
soft key menu to call up
Power
soft key menu as the right menu shown in