Typical Applications
16
Figure 3-7 Soft Key Menu of SOLT 2-Port Reflection Calibration Completion
There is no regulated sequence for open circuit, short-circuit, and loading
in practical calibration. You may choose arbitrarily.
Connect 3.5mm THRU connector to between Port 1 and Port 2. Press
[Transmission]
in
Full 2-
Port
soft key menu. Transmission soft key menu will pop up as shown in Figure 3-8. You may perform
calibration by pressing
[Unknown THRU]
as shown in the following picture as cables are not directly
connected but inserted with 3.5mm THRU connector in the middle. Transmission calibration is completed
when menu items turn to
[Unknown THRU]
.
Press
[DONE Transmission]
.
Figure 3-8 Soft Key Menu of Transmission Calibration Completion
Reflection
and
Transmission
calibrations are completed as shown in the following picture. Press
[Done FULL 2-Port]
to complete SOLT 2-port calibration.