Typical Applications
30
Figure 3-32 Modify TRL Calibration Class Assignments Dialog Box
As shown in the left figure of the following picture, on the
Calibration Kit Class
area, select
[TRL THRU]
, then select
[3 THRU 3.5mm THRU]
in the area of
Unselected Standards
. Then click the
button on the red circle to add the standard to the area of
Selected Standards
, the completion is as shown
in the right figure of the following picture.
Figure 3-33 Edit TRL THRU Class
In the same way, select the calibration standard for TRL reflection, as shown in the following
picture, select
[TRL REFLECT]
for calibration kit class, then select the
[1 OPEN 3.5mm Female]
option
in the area of unselected standards, then click the button on the red circle to add the standard to the area of
selected standards, the completion is as shown in the right figure of the following picture.