Typical Applications
29
Figure 3-30 Edit THRU Cal Standard
As shown in the following picture, click
[LINE]
and then
[OK]
,
Thru/Line/Adapter
dialog box
pops up. Input 0.48 to Delay Characteristics dialog box, then press
【
G/n
】
key of ENTRY on the front
panel to set the delay as 480ps. After setting, click
[OK]
to close the dialog box, thus LINE standard is
defined.
Figure 3-31 Edit Calibration Standard for Transmission LINE
In the
Class Assignments
area of
Edit Kit
dialog box, select
[TRL]
from the sub-menu, then click
[Edit…]
key and
Modify
TRL Calibration Class Assignments
dialog box pops up to set the newly-added
calibration standards as per class. Click
[LINE Z0]
in the
Calibration Reference Z0
area, then click
[THRU STANDARD]
in the
Testport Reference Plane
area, the completion is as shown below.