Typical Applications
13
Figure 3-2 Sweep Soft Key Menu
Press
【
Avg
】
key on front panel and press [
IF Bandwidth
] in soft key menu displayed on the
right side of display screen. press
【
10
】
→
【
/off
】
in the entry bar on screen to set IF frequency 10Hz.
Figure 3-3 Average Soft Key Menu
3.1.2
Calibration
To improve the accuracy of measurement results, you may calibrate the analyzer before measuring the
DUT to eliminate system errors. It is required to perform dual-port SOLT calibration for the measurement
of a filter. Calibration kit 20202(3.5mm) is selected suitable for the measurement cable.
Select calibration kit: press
【
Cal
】
key on front panel to call up soft key menu as shown on the
Summary of Contents for 3656 Series