Typical Applications
12
3.1 Filter Test
This section describes how to measure the characteristics of a 2GHz center frequency as an example, such
as insertion losses, out-of-band rejection, return loss, and etc.
3.1.1
Set up measurement Conditions
Press
【
Preset
】
on front panel to return the analyzer to the initial setup state, when it reaches
regulated pre-heating time. Default trace parameter is S11. The number of default sweep points is 201.
Default trace display scale is “10dB” and reference value is “0dB”.
Press
【
Start
】
and
【
Stop
】
on front panel to call up the following soft key menu as shown in the
following picture. Click
[Start]
, press
【
1
】
→
【
G/n
】
on the entry bar of the screen, and set the start
frequency 1 GHz; Click
[Stop]
,press
【
3
】
→
【
G/n
】
on the entry bar, and set the stop frequency 3 GHz.
Figure 3-1 Frequency Soft Key Menu
Press
【
Sweep
】
key on front panel to call up soft key menu as shown in the following picture.
Press [
Power
]
→
[
Power Level
], press
【
10
】
→
【
/off
】
on the entry bar, and set power level 10dBm .