Typical Applications
26
Figure 3-25. Press
[Start Power]
, press
【
+/-
】
→
【
15
】
→
【
/off
】
in input box in the upper right side
of screen and set up the start power as -15dBm. Press
[Stop Power]
, press
【
10
】
→
【
/off
】
and set up
the stop power as 10dBm. Press
【
Start
】
key on front panel. Press
【
3
】
→
【
G/n
】
in dot frequency
dialog box. Power sweep testing curves of amplifier are shown as the picture below. Press
【
Marker
】
key
on front panel and press
[Marker 1]
in soft key menu called up. press
【
+/-
】
→
【
15
】
→
【
/off
】
in
dialog box of Marker 1, Press
[Marker 2]
and activate Marker 2. Use adjust knob to move Marker 2. Then
you may find 1dB compression point of magnifier by comparing values of makers.
Figure 3-25 Power Sweep Test Curve
3.3 Customization of TRL Cal Kit
Apart from the basic functions of measuring the normal DUT, 3656 Series Vector Network Analyzer can
also provide users with accurate TRL calibration and measurement, particularly suitable for the
measurement under non-coaxial environment, which provides a more accurate solution for users to perform
measurement with non-standard coaxial cable. TRL calibration kit is mainly customized by the users,
therefore the analyzer supports user to modify the calibration kit definition or create customized calibration
kit.
This section introduces the calibration kit customization function of the analyzer, taking the definition of a
set of TRL calibration kit as an example. The TRL standard kits used in this example include OPEN,
THRU, and transmission LINE, with frequency range of transmission LINE being 200MHz-800MHz,
transmission LINE delay being 480ps, OPEN and THRU kits defaulted to be the ideal calibration kits.
Connect the USB keyboard and mouse to the DUT before the test, and create a TRL calibration kit as
follows:
Summary of Contents for 3656 Series
Page 1: ...1 3656 Series Vector Network Analyzer Quick Start Guide ...
Page 2: ......
Page 5: ......
Page 6: ......
Page 8: ......