Typical Applications
15
Press
[Reflection]
in
Solt 2-port
soft key menu in the right side of the above picture. Reflection
calibration menu will be shown in the following picture.
Figure 3-6 Solt 2-Port Reflection calibration
Connect Port 1 and Port 2 of the analyzer respectively with N-3.5mm cable. Connect 3.5mm
female OPEN standard to the other end of the cable connected to the PORT 1. Press
[PORT1 OPEN]
and
start to calibrate as shown in Figure 3-6. PORT1 open calibration is completed when menu item turns to be
[PORT1 OPEN]
.
In the same way, complete calibrations separately by connecting cable of Port 1 with
female SHORT/LOAD standards.
Connect cable of Port 2 of the analyzer with female OPEN calibration kit. Press
[PORT2 OPEN]
and start to calibrate as shown in Figure 3-6. PORT2 open calibration is completed when menu items turn
to be
[PORT2 OPEN].
Take the same steps and complete calibration separately by connecting cable with
female SHORT calibration kit and female LOAD calibration kit. The soft key menu status after reflection
calibrations of two ports are completed is as shown in the following picture. Press
[DONE Reflection]
.