Typical Applications
33
the set limit line value before PASS. Users may set up multi-segment limits as required, the first segment of
limit line in this example is set as follows:
[Type]
:MAX
; [Start Stimulus]
: 1.0GHz,
[Stop Stimulus]
:
1.7GHz;
[Start Response]
: -120dB,
[Stop Response]
: -120dB;
[Limit Type]
: SLOP.
Figure 3-38 Limit Table Setting
The second segment of limit line is set as:
[Type]
:MAX
; [Start Stimulus]
: 2.35GHz,
[Stop
Stimulus]
: 2.6GHz;
[Start Response]
: -110dB,
[Stop Response]
: -110dB;
[Limit Type]
: SLOP. The third
segment of limit line is set as:
[Type]
:
MIN; [Start Stimulus]
: 2.01GHz,
[Stop stimulus]
: 2.15GHz;
[Start
Response]
: -3dB,
[Stop Response]
: -3dB;
[Limit Type]
: SLOP. As shown in the below figure, if the test
result meets the pre-set limit value, then
Trl: Pass
will display on the lower right corner of the screen,
indicating pass the test.
Figure 3-39 Pass Limit Test