Typical Applications
20
multiple traces. Steps are as follows: Press
【
TRACE
】
key on front panel and press
[New S-Parameter]
in soft key menu. Four S parameters will be displayed. You may select
[S11], [S12],
and
[S22]
in sequence.
Figure 3-15 New S Parameter Soft Key Menu
Press
【
Trace Prev
】
key and activate S12 trace. Press
【
Scale
】
key on front panel. Press
[Scale]
in scale soft key menu on the right side of screen. Press
【
20
】
→
【
/off
】
in the entry box on the screen,
set scale 20dB; press
[Reference Level]
and press
【
+/-
】
→
【
20
】
→
【
/off
】
in the entry box on the
screen, set Ref value -60dB . Curves of four S parameters of filter test are shown as the following picture.
Figure 3-16 Filter Test Diagram of 4 S Parameters
The analyzer provides function of measurement result output after test is completed. You may
output and save all calibration data and measurement status of the analyzer. Press
【
Save
】
key on front
panel and soft key menu on the right side of screen will be as shown in the following picture. [
Store