Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
5-1
5
Troubleshooting and
Maintenance
Contents
◆
Self-tests 5-1
◆
Indicator troubleshooting table
5-3
◆
Scheduled maintenance
5-4
◆
Authorized repair service
5-6
◆
Frequently Asked Questions
5-7
This section presents information about the AED diagnostics
self-tests, maintenance, and service indications.
Self-tests
The AED has a comprehensive self-test system that automatically
tests the electronics, battery, pads, and high voltage circuitry.
Self-tests are also activated every time you open and close the
AED lid.
When performing the self-tests, the AED completes the following
steps automatically:
1.
Turns itself on, and the Status Indicator changes to red
2.
Performs the self-test
3.
If successful, the Status Indicator reverts to green
4.
Turns itself off if the lid is closed.
There are three types of automatic self-tests:
◆
The daily self-test checks the battery, pads, and the electronic
components