6-10
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Attenuated energy values with Cardiac Science
pediatric electrodes and STAR® biphasic waveform
The waveform generated by the AED is a Biphasic Truncated
Exponential waveform. The following is a graph of the waveform
voltage as a function of time when the AED is connected to a 50
Ohm resistive load using pediatric pads.
The Biphasic Truncated Exponential (BTE) waveform uses variable
energy. The actual energy delivered will vary with the patient’s
impedance and the device will deliver a shock when impedance
is between 25-180 Ohms. Energy will be delivered at three
different levels referred to as ultra-low variable energy, low
variable energy, and high variable energy as shown in the
waveform tables on the following pages.
Typical Pediatric Waveform: Low Energy (200 VE) 50 Ohm Patient Impedance
Technical Data
2000.0
1500.0
1000.0
500.0
0.0
-500.0
-1000.0
-1500.0
-1.000
0.000
1.000
2.000
3.000
4.000
5.000
6.000
7.000
8.000
9.000
10.000
TIME (ms)
VOL
TS
High Energy Waveform with 50 Ohm Resistive Load
High Variable Energy / 50 Ohm
- 600
- 400
- 200
0
200
400
600
800
1000
3
6
9
12