1-12
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Guidance and manufacturer’s declaration—
electromagnetic immunity
The AED is intended for use in the electromagnetic environment specified below. The
customer or the user of the AED should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment—guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidi-
ty should be at least 30%
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
80 A/m
Power frequency magnetic
fields should be at levels no
higher than those character-
istic of a typical location in
typical heavy industrial and
power plants and the control
rooms of H.V. substations.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Conducted RF
3 Vrms
Not
Applicable
IEC 61000-4-6
150 kHz
to 80 MHz
outside
ISM bands
a
10 Vrms
150 kHz to
80 MHz in
ISM bands
a
Not
Applicable
Product Information and Safety