3-12
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Table 3-8: Other prompts
Voice Prompt
Text display
Situation
“Battery low”
BATTERY LOW
Occurs once when the battery voltage
becomes low, although a rescue can
continue for approximately 9 more
shocks. When the battery is too low
to do a rescue, the following will occur:
BATTERY LOW shows on the LCD
Smartgauge™ battery status indicator
turns red.
AED beeps once every 30 seconds
while the lid is closed
You must replace the battery before
continuing with the rescue. If
completely depleted, all AED activity
will terminate.
“Analysis interrupted. ANALYSIS INTERRUPTED When the AED detects ECG noise
Stop patient motion.” STOP PATIENT MOTION
artifact, stop moving or touching
the patient. Remove other electronic
devices within a 5 meter radius.
“Open lid to continue OPEN LID
When the lid is inadvertently closed
rescue.”
TO CONTINUE RESCUE
during a rescue, this prompt will
repeat for 15 seconds.
Rhythm changed.
RHYTHM CHANGED
When the device is prepared to
Shock cancelled.”
SHOCK CANCELLED
shock then detects a change in
rhythm and therefore
cancels the shock.
“Remove cable to
REMOVE CABLE
When a serial communication cable
continue rescue.”
TO CONTINUE RESCUE
is connected to the AED during a
rescue, the phrase repeats until the
cable is disconnected.
“Communications
COMMUNICATIONS MODE When the lid is open and the serial
mode”
communication cable is plugged
into the AED.
“Service required”
SERVICE REQUIRED
Occurs after the self-tests determine
that the AED is not functioning
properly. The prompt “SERVICE
REQUIRED” will be heard when the
lid is opened. The red SERVICE
indicator will illuminate. After closing
the lid, an indicator beep will be
heard until the battery is removed
or becomes completely depleted.
Getting Started