Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
6-5
Table 6-1: Parameters (continued)
Parameter Detail
Immunity EM
IEC 61000-4-3, Level X, (20V/m)
IEC 60601-2-4 (20V/m)
Magnetic
IEC 61000-4-8
IEC 60601-2-4
ESD
IEC 61000-4-2
IEC 60601-2-4
6kV contact discharge, 8KV air gap discharge
Environmental Conditions Free Fall Drop: 1 meter per 60068-2-31:2009, Environmental
testing – Part 2-31: Tests - Test Ec: Rough handling shocks,
primarily for equipment-type specimens.
Bump: IEC 60068-2-27:2011, Environmental testing –
Part 2-27: Tests Ea and guidance: Shock.
Vibration (Random): IEC 60068-2-64:2008, Environmental
testing – Part 2-64: Tests – Test Fh: Vibration, broadband
random and guidance.
Vibration (Sine): IEC 60068-2-6:2007, Environmental testing
– Part 2-6: Tests – Test Fc: Vibration (sinusoidal).
Enclosure Protection: IEC 60529, IP24
Vibration (random): RTCA DO-160G:2010, Section 8
Airborne use: RTCA DO-160G Sec 8, category U-Airplanes
and Helicopters
Shipping and
ISTA Procedure 2A
Transportation Conditions
Technical Data