Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
1-5
WARNING! Reduced therapy delivery.
Failure to remove blue liner completely could impact therapy delivery.
Caution. Short-term use only. Not for pacing.
DO NOT open defibrillation pads package until ready to use.
Short term use only.
Pads are not intended for use in pacing.
Caution. Equipment Damage.
Do not pull on the lead wire to separate the pads from the blue liner.
Note: Store pads at room temperature.
Note: Pads are intended for adult use.
WARNING! Equipment not functioning.
If the AED stops functioning during a rescue, continue to perform
CPR as needed until EMS personnel arrive.
WARNING! Possible Radio Frequency (RF) Susceptibility
Do not use the AED in locations where large electromagnetic or
RF fields can be expected to occur.
Refer to Chapter 1: Electromagnetic Emissions Standards Compliance
for additional information.
WARNING! Possible Interference with Implanted Pacemaker
Therapy should not be delayed for patients with implanted pace-
makers and a defibrillation attempt should be made if the patient
is unconscious and not breathing. The AED has pacemaker de-
tection and rejection. However, with some pacemakers the AED
may not advise a defibrillation shock. (Cummins, R., ed., Advanced
Cardiac Life Support; AHA (1994): Ch. 4)
When placing pads:
•
Do not place the pads directly over an implanted device.
•
Place the pad at least one inch from any implanted device.
WARNING! Electromagnetic Compatibility
Use of accessories or cables other than those specified, with the
exception of accessories and cables sold by Cardiac Science
Corporation as replacement parts for internal components, may
result in increased emissions or decreased immunity of the AED.
Product Information and Safety
!
!
!
!
!
!
!