Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
6-9
Table 6-4: High Variable Energy Powerheart® G3 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Current*
(A)
Voltage*
(Volts)
Duration*
(MS)
Current*
(A)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
75
1869
3.3
40
997
3.2
305
50
38
1906
4.5
24
1220
3.2
270
75
26
1918
5.8
17
1306
3.2
246
100
19
1925
7
14
1351
3.2
229
125
15
1928
8.3
11
1378
3.2
216
150
13
1931
9.5
9
1396
3.2
207
175
11
1933
10.8
8
1408
3.2
200
* All values are typical.
**Actual energy del/- 15%.
Technical Data