Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
5-5
3.
Check the expiration date on the pads.
4.
Check the Smartgauge™ on the front panel to ensure the
battery has sufficient power. Replace the battery if the
indicator is RED.
5.
Listen for the voice prompts. Additionally, check that the
display shows text prompts that correspond to the audio.
6.
Close the lid and verify that the STATUS INDICATOR changes
from GREEN to RED during its power down self-test. After
approximately 5 seconds, verify that the status indicator
returns to GREEN.
Annual maintenance
Perform the following tests annually to confirm that the
diagnostics are functioning properly and to verify the integrity of
the case.
Check the integrity of the pads and circuitry:
1.
Open the AED lid.
2.
Remove the pads.
3.
Close the lid.
4.
Confirm that the STATUS INDICATOR turns RED.
5.
Open the lid and confirm that the Pads indicator is lit.
6.
Reconnect the pads and close the lid.
7.
Make sure the expiration date is visible through the clear
window of the lid.
8.
Check to make sure that the STATUS INDICATOR is GREEN.
If the pads are not installed properly, the Pads indicator will
illuminate. Contact Cardiac Science Technical Support
(see
Contact information
on page 1-2) or outside the U.S.,
your local Cardiac Science representative.
9.
Open the lid and confirm that no diagnostic indicators are lit.
10.
Check the expiration date of the pads; if expired, replace
them.
11.
Check the pads packaging integrity.
12.
Close the lid.
Troubleshooting and Maintenance