6-2
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Table 6-1: Parameters (continued)
Parameter Detail
Dimensions
Height: 3.3 in (8 cm)
Width: 10.6 in (27 cm)
Depth: 12.4 in (31 cm)
Weight (Batteries and Pads) 6.6 lb (3.10 kg)
Operating and Standby
Temperature: 32°F to 122°F (0°C to 50°C)
Environmental Conditions Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Pads
Self-adhesive, disposable defibrillation pads
Minimum combined surface area: 35.3 in
2
(228 cm
2)
Extended length of lead wire: 4.27 ft (1.3 m)
9146 Lithium Battery
Output voltage: 12VDC
Specifications
Batteries are non-rechargeable
Lithium content: .32 oz (9.2 g)
Check local regulations for disposal information
Estimated Shelf Life (from date of manufacture): 5 Years
Typical Shocks: 290 shocks
Note:
The battery operating life depends on the type of
battery, device settings, actual usage, and environmental
factors. Battery was tested with a G3 AED device with
Standard prompt set and CPR set to 60 seconds.
Storage and Transport
Technical Data
Configuration
Transport Storage
Use
Packaged System
(packaging, unit,
pads, battery)
5 days at
-30°C to
+65°C
2 years
at 0-50°C
(pad life)
N/A
Unpackaged
System without
accessories
5 days at
-30°C to
+65°C
10 years
at 0-50°C
N/A
Unpackaged System
with Accessories
(batteries and pads)
5 days at
-30°C to
+65°C
2 years
at 0-50°C
(pad life)
2 years
at 0-50°C
(pad life)
Pads (packaged)
5 days at
-30°C to
+65°C
2 years
at 0-50°C
(pad life)
N/A
Battery (packaged
or unpackaged)
5 days at
-30°C to
+65°C
5 years at
20-30°C
4 years at
0-50°C