Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
3-3
Diagnostic panel
The diagnostic panel has the following indicators:
1.
Smartgauge™ battery indicator
2.
Pads indicator
3.
Service indicator
4.
Shock Button (Powerheart® G3 Elite Semi-Automatic model
9790E only)
Smartgauge™ battery status indicator
The Smartgauge™ Battery Status Indicator
has five LEDs, four green and one red.
The right four green LEDs display the
remaining capacity of the battery much like
a fuel gauge. With use, the green LEDs gradually go out, from right
to left, as battery capacity decreases. When the green LEDs go out
and the red LED lights up, replace the battery.
Note:
When the red LED initially lights up–upon lid opening or at
any time during a rescue–a BATTERY LOW prompt will be issued
at once. However, the AED is capable of delivering at least 9
defibrillation shocks after the first BATTERY LOW prompt is issued.
When the AED battery cannot deliver any more shocks, the AED
shows BATTERY LOW on the text display, and the red battery LED
illuminates. To continue the rescue, leave the lid open, remove the
battery, and replace with a fresh battery. If battery replacement
takes longer than 60 seconds, the first rescue will be terminated
and a second rescue will begin upon insertion of battery.
Note:
When the battery is depleted, neither the LED nor the text
display illuminates.
0%
100%
1
2 3
4
0%
100%
Getting Started