5-2
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
◆
The weekly self-test completes a partial charge of the high
voltage electronics in addition to the items tested in the
daily self-test
◆
During the monthly self-test, the high voltage electronics
are charged to full energy in addition to the items tested in
the daily self-test.
In addition, self-tests will be initiated upon opening the lid and
again upon closing the lid.
If the self-test detects an error, the Status Indicator remains red.
Upon closing the lid, an audible alert will be issued. The diagnostic
panel under the lid indicates the source of the problem according
to Table 5-1 on page 5-3.
Troubleshooting and Maintenance