3-6
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Getting Started
Setting the AED internal clock
For US models, the internal clock is preset to Central Standard
Time. You can reset it to your local date and time. To set the clock,
you need a Windows 7 or newer PC, RescueLink software
installed, and the AED serial cable connected to the PC.
To set the clock:
1.
Ensure that the PC is set at the correct local time and date.
2.
Open the lid of the AED and run the RescueLink software on
the PC.
3.
Connect the cable to the serial port on the AED.
4.
Verify that the voice prompt states “Communications Mode”.
5.
Click Communications on the main menu. Select AED Date
and Time.
6.
Click on the Get button to review the current time in the AED.
7.
If the time and date are incorrect, click Set to set new time
and date. The AED date and time will automatically be
updated to the PC’s time and date.