Environment Specifications
OMNERA™400T
0180-095-485
PlanningGuide
Page 2-1
Environment Specifications
Placement of Non-medical Components
Non-medical components may not be placed inside the patient environment according to reg-
ulations and standards (IEC 60601-1).
The standard defines a sphere of 1.5 m around the patient as the patient environment. The
Ceiling support can be operated across the entire rail grid, i.e. the 1.5 m boundary must be
placed outside the rail grid. Also take into account the patient entries and exits and the possi-
bility for the patient to walk around in the examination room. In practice, this means that no
non-medical components can be placed inside the walls of the examination room.
Image System Computer
Image system computer used and other similar devices are normally non-medical compo-
nents. The image system computer and the including touch-screen used in the System is no
exception and may therefore not be placed inside the examination room.
Room and Installation Specifications
Wall Clearance
According to regulations and standards (IEC 60601-1), a minimum distance of 500 mm be-
tween moving motorized parts and surrounding walls, and/or permanently installed objects
must be considered when planning the room layout. Since the Ceiling suspended x-ray tube
support and the Wallstand is manually operated (except from the Z-movement of the Ceiling
suspended x-ray tube support), the clearance requirement does not apply to Table.
The table top movement can be limited to allow a smaller room size, or if objects are placed
within the 500 mm zone. The stroke is adjustable via mechanical end stops, placed on each
side of the table top.
Ceiling Height
The System has three configurations allowing a ceiling height between 2500 mm and 3050
mm.
Table 2-1
Ceiling height
(mm)
Vertical column
stroke (mm)
Ceiling rails
2500-2700
1450
Low ceiling rails.
2700-2850
1450
Standard ceiling rails. (Standard con-
figuration).
2850-3050
1700
Standard ceiling rails.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...