![Canon OMNERA 400T Planning Manual Download Page 54](http://html1.mh-extra.com/html/canon/omnera-400t/omnera-400t_planning-manual_4018787054.webp)
Room Layout
Planning Guide
0180-095-485
OMNERA™400T
Page 5-6
Room Layout 2b
Figure 5-3
This layout is also a variation of layout 1. It is only valid with a tilting Wallstand and this option
may be considered if the room layout, e.g. placement of doors, does not allow layout 1 to be
chosen. The downside of selecting this layout is the larger room that is required and the ex-
tended amount of mass that has to be handled during examinations using the Wallstand. It is
also limited when doing lateral projections and positioning over tilted detector (Pos. A).
Focal spot coverage, Beta pos B
Orientation point
Focal spot coverage, Beta pos A
X
Y
8
6
9
974
2252
3
1
1
3
2
5
0
(*
9
)
4200
5
1
5
0
2
5
0
(*
9
)
768 (*2)
428 (*2)
4
8
5
(
*4
)
1
1
9
0
(
*3
)
1
5
7
5
2100 (*1x)
Pos A
Back
Front
H
e
a
Current Wall stand setup: Tilt, left loaded
1100 (1100-2980LL) (*5)
(1220-3351RL) (*5)
180
1
6
7
5
(
*1
y
)
(*7)
1220(1220-3101LL) (*5)
(849-2981RL) (*5)
Pos B
4
1
3
(
*6
)
4
1
3
(
*6
)
125
4
9
2
452 (452-2333LL) (*8)
(81-2213RL) (*8)
672 (672-2552LL) (*8)
(792-2923RL) (*8)
Pos D
1654
in
Orientation po t
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...