Room Layout
Planning Guide
0180-095-485
OMNERA™400T
Page 5-12
Room Layout 5
Figure 5-6
This layout should be used if the System configuration consists of a Ceiling suspended unit,
Wallstand with tilt and a movable Table. The ceiling rail of the Ceiling suspended unit must be
reversed to enable the tube and the detector/cassette to be aligned, when the image receptor
is placed under the movable table top. The detector holder has a left loading configuration in
this layout
For cable length see
System Connections on page 18.
Foc
a
l spot cover
ag
e, Bet
a
pos B
Foc
a
l spot cover
ag
e, Bet
a
pos
A
X
Y
7
94
22
5
2
3500
4
5
0
0
2
48 (*
2
)
2
5
0
(*
4
)
Pos
A
Current W
a
ll st
a
nd setup: Tilt le
f
t lo
a
ded
(*
6
)
Pos B
2
50 (*1)
1
2
5
Pos D
2
4
1
8
0
2
2
5
5
(*
5
)
1
5
7
5
300
2
(*3)
4
5
4
4
9
2
3
1
1
3
M
o
v
a
b
le
t
a
b
le
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...