Room Layout
Planning Guide
0180-095-485
OMNERA™400T
Page 5-8
Room Layout 3
Figure 5-4
In this layout (and in layout 4) is the ceiling rail of the Ceiling suspended unit, reversed which
allows the System to operate in a long and narrow room. However, by reversing the ceiling rails
the ergonomics is not optimized since the operator has to move more mass during positioning.
This layout may be considered if the ceiling preparation does not allow the previous options to
be selected or if the size of the room is more suitable.
For cable length see
System Connections in chapter 5.
Focal spot coverage, Beta pos B
Orientation point
Focal spot coverage, Beta pos A
X
Y
4
5
6
794
2252
3
1
1
3
3
3
0
(*
9
)
3500
4
8
0
0
2
5
0
(*
9
)
248 (*4)
5
4
8
(
*2
)
1673 (*1x)
Pos A
B
a
c
k
F
ro
n
t
Head
Orientation point
Current Wall stand setup: Tilt left loaded in lower corner.
660 (660-830RL) (*5)
2100
(
*1
y
)
(*7)
Pos B
8
0
(
*8
)
125
4
9
2
Pos D
2
5
0
(2
*)
1425 (*3)
180
2
4
660 (660-830LL) (*5)
412 (412-582RL) (*6)
412 (412-582LL) (*6)
3
6
0
2
Tilt right loaded in upper corner.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...