Environment Specifications
Planning Guide
0180-095-485
OMNERA™400T
Page 2-10
Wall Specifications
The walls are normally only used for mounting the cable outlets for the Ceiling suspended x-
ray support, the Wallstand and the System cabinet. Therefore, the only specification for the
walls, is that it must be possible to mount and secure the cable outlets. Screws, bolts or similar
for mounting the cable outlets, must be supplied (not included in shipment). The regulations
and specifications on page 2-1 must be followed.
The Installation cube shall be mounted to the wall. There are four wall fixation points, and each
fixation point must withstand a load of at least 0.5 kN. The wall attachment allows to place the
installation cube 0-1000 mm from the wall. If the distance to the wall is small, the attachment
must be shortened to a correct length. When the Installation cube is placed close to the wall,
make sure that a suitable placement of the cable outlet is possible.
Figure 2-7
750
750
750
750
Minimum distance, 0 mm
Maximum distance, 1000 mm
0.5kN
0.5kN
0.5kN
0.5kN
76
86
4
0
22
10.5 (2x)
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...