![Canon OMNERA 400T Planning Manual Download Page 72](http://html1.mh-extra.com/html/canon/omnera-400t/omnera-400t_planning-manual_4018787072.webp)
Component Specification
Planning Guide
0180-095-485
OMNERA™400T
Page 6-4
Vertical Movement Range
To be able to accomodate three different ceiling heights, one short and one long column are
used. The shorter column is used in configuration 1 and 2 whereas the longer column is used
in configuration 3.
Fully extracted, the short column reaches a length of 2180 mm and the longer column reaches
a length of 2530 mm. Adding the built-in measurements from the suspension, gives the total
length of the columns. The longer column (used in configuration 3) is also 100 mm longer when
fully retracted, giving a minimum length of 830 mm (short column, 730 mm).
Considering these facts, a trade-off between a high and low elevation must always be consid-
ered. For more information about the configurations,
see page 2-1.
NOTE
!
Configuration 1 is always delivered with a 2748 mm traverse (Low ceiling), limiting the move-
ment range of the ceiling wagon.
Figure 6-2
170
0
145
0
1450
877
816
71
1
2500 -
2700
2700 -
2
850
2850 -
3
050
Configuration
1
Configuration
2
Configuration
3
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...