System
Electrical Characteristics
Environmental Requirements
Ceiling Suspended X-ray Tube Support
General
Configuration
Weight
Mains voltage for the System
400V 3N, 50-60 Hz
400V 3~
480V 3~
150A (Short term peak value), recom-
mended fuse 63 A thermal breaker).
Ambient transport and storage temperature -40
o
C - +70
o
Ambient operating temperature
+10
o
C - +40
o
Transport and storage humidity (relative)
10-90%, non-condensing
Operating humidity (relative)
30-75%, non-condensing
Maximum transport and storage altitude
500-1060 hPa
Maximum operating altitude
700-1060 hPa
Rotation range ceiling (beta)
-180
o
(-5
o
) - >+163
o
Rotation range tube arm (alpha)
-180
o
(-5
o
) - >+163
o
Column (Z stroke)
1700 mm, 1450 mm
Heat dispation
689 BTU/hr
Ceiling suspended x-ray tube support
The Ceiling suspended x-ray tube support is
a mechanical part of an X-ray system.
Ceiling suspended x-ray tube support
127 kg
Tube and Collimator
40 kg, maximum allowed weight
Traverse rail X
60 kg
Ceiling rail Y (4m standard)
16 kg
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...