Room Layout
Planning Guide
0180-095-485
OMNERA™400T
Page 5-16
Room Layout 7
Figure 5-8
Layout 7 is designed for very small rooms with a low ceiling height (>=2500). In this layout the
Y rails of low ceiling rail type are mounted directly to the ceiling. To minimize the required room
space, the table top could be fixed in the smallest room (in longitudinal direction) and the sys-
tem cabinet is not placed within the room.
This layout enables the same possibilities as the other options but with some restrictions and
limitations.If the table top is fixed, a full patient coverage is not possible without moving the pa-
tient.
Lateral projections from behind the Table are also restricted due to the limited space available.
Focal spot coverage, Beta pos B
Orientation point
Focal spot coverage, Beta pos A
X
Y
659
1945
3
1
1
3
3000
4
0
0
0
Pos A
B
a
c
k
F
ro
n
t
Head
Orientation point
Current Wall stand setup: Tilt right loaded
2100
(
*2
y
)
Pos B
125
4
9
2
Pos D
250 (*1)
750 (*1)
2
5
0
(
*1
)
125 (*3)
1100 (*2x)
8
0
(
*3
)
S
ID
>
2
7
0
0
2
5
0
(
*1
)
3
3
0
(
*3
)
2
0
0
2930
4
0
0
0
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...