Environment Specifications
OMNERA™400T
0180-095-485
PlanningGuide
Page 2-11
Electrical Building Installation
The Machinery Directive 2006/42/EC requires the System to be fitted with means to isolate it
from all energy sources.
NOTE
!
When Service or Maintenance will be performed the technician shall lock the equipment from
all energy sources.
Power Ratings and Line Requirements
The System requires a three-phase electrical line with a protective earth ground (4 or 5 wires).
The transformers in the system cabinet requires a tap configuration (*1), the generator has an
automatic main line selection (no transformer tap configuration required). A disconnecting de-
vice shall be incorporated external to the equipment according to the national wiring rules.
•
3 Phase VAC ±10%
•
400VAC 50Hz
•
400VAC with neutral 50Hz
•
480VAC 60Hz
•
Maximum wire gauge 4 AWG (25mm²)
•
Required fuse 63A, 3-phase, b-curve
NOTE
!
*1 For more information about tap configuration see Installation and Service Instructions.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...