Room Layout
OMNERA™400T
0180-095-485
Planning Guide
Page 5-1
Room Layout
Installation Proposal
The layout proposals, displayed in the drawings on the next pages, are chosen to optimize the
performance of the System with usability and ergonomics in mind. The chosen layout proposal
considers the tube heel-effect, cable conduits, work space etc, as well as safety issues such
as wall clearance and possible collision zones. The room appearance (furniture, lights, fire pro-
tection devices etc.) is not considered in the layouts.
The versatility of the System allows a minimum room dimension of 3500 x 4500 mm. Imple-
menting the System in a small room (<4200 x 4500) limits the moving range of the table top
(see wall clearance in chapter 2). The 4200 x 4500 mm room dimension allows an 800 mm
table top stroke which gives total body coverage of a 2500 mm patient. The 800 mm table top
stroke is used in all layout proposals, except for layout 6 and 7 where the table top is fixated in
the longitudinal direction.
The layouts are valid for the tube Varian Rad 14/Diamond and collimator Ralco 302 (manual).
For other types of tubes and collimators when planning the room, see page 6-2.
Implementing the System in a small room, limits the moving range of the table top.
Reference
Recommended placement of doors
Alternative placement of components
Movement range
Orientation point, correspond to the mounting point E of the Table. For more information see
Floor specification
for the Table
,
in chapter
Environment specification
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...