Room Layout
OMNERA™400T
0180-095-485
Planning Guide
Page 5-15
Lateral projections from behind the Table are also restricted due to the limited space available.
The table top must be pushed to the front to be able to position the tube for a lateral projection,
but even then the position flexibility is restricted.
The detector holder has a right loading configuration in this layout.
For cable length see
System Connections on page 18.
Wallstand loading and position should be carefully planned, since it will have large impact on
the final layout of the room. If the use of armrest is restricted to one side it is also possible to
move Wallstand closer to wall or Table.
If recommended dimensions are overridden, necessary notifications, warnings and instruc-
tions have to be considered. This includes safety distances and trapping zones around motor-
ized movements according to IEC 60601-1 3rd edition.
•
(*1)
must not be less than 30 mm to be able to install the cube.
•
(*2y)
is showing the Table centered in Y-direction. It makes it possible to cover the
Table and horizontal Wallstand when beta is in position B.
(*2x)
has to be 1100 to
make it possible to have full lateral projection functionality. Still the table top has to
be in front position. Working area of Wallstand has to be considered when moving
Table in X-direction.
•
(*3)
enables Ceiling stand to project in the center of the tilted detector holder.
•
Dimensions state relation between components and have to be adapted to the
planned site.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...