![Canon OMNERA 400T Planning Manual Download Page 15](http://html1.mh-extra.com/html/canon/omnera-400t/omnera-400t_planning-manual_4018787015.webp)
Environment Specifications
OMNERA™400T
0180-095-485
PlanningGuide
Page 2-7
Floor Specifications
The Table and the Wallstand must be installed on a solid base with sufficient load capacity (not
within scope of this document), e.g. concrete.
The floor must be able to withstand the pull forces supplied on the expansion bolts. The pull
forces for the components are listed in Table 2-4, see also
Fig. 2-4
and
Fig. 2-5
.
The Table and the Wallstand must be leveled to 1 mm/m.
Figure 2-4 Drilling template,Table
Figure 2-5 Marking template, Wallstand (Isolation Plate)
A
D
B
C
B
C
D
A