Room Layout
OMNERA™400T
0180-095-485
Planning Guide
Page 5-17
The table top must be pushed to the front to be able to position the tube for a lateral projection,
but even then the position flexibility is restricted.
The detector holder has a right-loading configuration in this layout.
For cable length, see
System Connections on page 18
.
If recommended dimensions are overridden, necessary notifications, warnings and instruc-
tions have to be considered. This includes safety distances and trapping zones around motor-
ized movements according to IEC 60601-1 3rd edition.
•
(*1)
is just reference dimensions, if less than 100 mm, special arrangements regard-
ing ceil attachment might be considered.
•
(*2y)
is showing the Table position in Y-direction. It makes it possible to cover the
Table and horizontal Wallstand when beta is in position B.
(*2x)
has to be 1100 to
make it possible to have full lateral projection functionality. Still the table top has to
be in front position. Working area of Wallstand has to be considered when moving
Table in X-direction.
•
(*3)
enables Ceiling stand to project in the center of the tilted detector holder when
beta is in position B.
•
Dimensions state relation between components and have to be adapted to the
planned site.
Summary of Contents for OMNERA 400T
Page 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Page 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Page 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Page 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Page 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Page 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Page 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Page 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Page 89: ...Intentionally left blank...