Installation and Servicing space
OMNERA™400T
0180-095-485
PlanningGuide
Page 3-15
Table
Front - Rear / Left - Right Direction
In the figure below, a lateral space required for: mounting/dismounting the table top is not con-
sidered.
When the table top is to be mounted/dismounted for installation and maintenance, a space of
2424 mm or more need to be reserved, either on the right or the left of the Table, as a longitu-
dinal space of the table top.
Figure 3-15
<Rear>
<Front>
<Right side>
f
Table top
Table top movable range
<Le
t side>
X
Y
1
50 +20/–10mm
2424 mm
500 +20/–10mm
500 +20/–10mm
150 +20/–10mm
808 mm