TCi Operator Manual
91
Insert the C-Therm TCi Software CD into the laptop and press the Open File button. Locate the
Authorization TH89-05-XXXXX.xml file on the CD and click the Add Keys button. The calibrations
will be available for use.
The cooling time in my test changed and is not what is set in the calibration method.
If the computer clock is changed while the sensor is cooling, the cooling time may be drastically
changed. Contact C-Therm customer technical support at 506 462-7204 or
to fix the problem.
My sensor is reading very erratic temperatures.
Abort the test in progress and turn off the power on the C-Therm TCi controller. Check all
connections and restart the controller. If the problem is not solved, examine the sensor chip for
any cracks or signs of damage. Contact C-Therm customer technical support at 506 462-7204 or
.
Summary of Contents for TCi