Due to variability between batches, there are multiple calibration material records for LAF 6720
and Phosphor Bronze. When performing a test, select the record matching the label on the
reference sample.
Step 9 – Click the Update button to check the ambient temperature (Optional).
Step 10 – Select Conductivity or Effusivity to display.
Note: Results for both properties are available at the end of the test.
Step 11 - Click the Get Sample button.
The result is displayed. All results should be within 5% of the predicted value (displayed under
the reference material name). If the result is in a red cell, the diagnostic has failed. If the
diagnostics fail, a new calibration may need to be performed. Contact C-Therm Technologies
Ltd. for troubleshooting.
Note: The following error is displayed if the Get Sample button is clicked before selecting
a reference material.
Note: The expected values for each calibration are calculated by the software. They are
also listed in Appendix D.
Step 12 - Click the Finish button.