TCi Operator Manual
60
Step 5 – Click the Add Row button once for each material in the group.
Step 6 – Set each row to one of the materials.
Step 7 – Click the Save button.
6.6
Create Calibration Method
Step 1 – Open the Calibration Method table.
Step 2 – Click the Add button.
Step 3 – Enter the Calibration Method name.
Step 4 – Select the Instrument Family.
Step 5 – Select the Calibration Material Group.
Step 6 – Click the Ignore Calibration Results from Chip checkbox.
Step 7 – Enter the Test Time in seconds.
Step 8 – Enter the Start Time in seconds.
Step 9 – Enter the Cooling Time in seconds.
Step 10 – Enter the Frequency. Note: If the test time is longer than 5 seconds, the
frequency must be changed to 50 Hz.
Step 11 – Select the Power Level. Use 90 for foams, liquids, powders, polymers, and
ceramics. Use 125 for metals.
Step 12 – Set the valid ranges (optional).
6.7
Acquire Calibration Data
Step 1 – Click User Calibration in the Tools menu.
Step 2 – Click the Acquire Data option from the User Calibration menu.